top of page
Recent Posts

276 Rapid Determination of Strontium in Water Using Empore Strontium Disk

This application note establishes a rapid analysis method for Strontiumin in water in emergency situations. The 1.00 L water sample was acidified and filtered with Empore Strontium solid-phase extraction disk, and then measured Strontium levels on the extraction disk with a low-background α/β counter. The method is simple to operate with stable chemical recovery rates. The lower detection limit of the method is 0.019 Bq/L under typical conditions, and the whole experimental process takes about 2 hours, which is suitable for the rapid determination of Strontiumin in water in emergency situations.


276 Empore Rapid Determination of 90Strontium in Water Using Empore Strontium Disk
.pdf
Download PDF • 246KB


Search By Markets
Search By Instruments

Leaders in GC Sample Introduction Technology

bottom of page