Heated Zone

Setpoint

Valve Compartment

Ambient to 200°C

Transfer Line(s)

Ambient to 260°C

Sample Tube Desorb

Ambient to 350°C

Sample Tube Idle

Ambient to 99°C

Focusing Trap Desorb

Ambient to 375°C

Focusing Trap Idle

Ambient to 99°C

Heating Accuracy

    All heated zones ± 2% of range minimum

Heating Rates

    Sample Tube: 1000°C/min. typical

    Focusing Trap: 900°C/min. typical

Timed Events

    A total of five sequence positions are designated: Ready (Idle), Tube Dry, Tube Heat, Trap Heat, and GC Time. Each event has a time (0 to 99.9 minutes in 0.1 minute increments) or off status (infinite).

Sampling Flow

  • Integral Porter® mass flow controller may be set from 0.00 to 2.55 slpm flow rate.
  • Accuracy & linearity: ±1% of full scale
  • Temperature Coefficient: 0.1%/°C of full scale
  • Pressure Coefficient: 0.1%/atmosphere
  • Repeatability: within 0.2% full scale at any constant temperature within operating temperature range
  • Response time: 1.2 seconds

Cycle Counter

    Range is1-7999 cycles, or off (infinite). System can be programmed to process a set number of samples, and then return to ready (idle) status.

Carrier Gas/Valving

  • Dual 8-port 1/8" stainless steel Valco® switching valves and electric actuators.
  • Sample transport tubing: sample path consists of 1/8" & 1/16" stainless steel tubing, silanized glass, and deactivated fused silica.
  • Transfer line: provision for two 1/4" O.D. heated jacket lines, one 36" heated jacket is supplied, internal surface is 0.25 mm I.D. fused silica.
  • Column flow: connections are provided for focusing trap desorb/column flow to be controlled by the EPC output from Agilent GCs. An optional flow control module is available for use if EPC is unavailable.
  • Glass tubing requirements: the sample collect chambers accept 6 mm and 10 mm (necked down to 6 mm) O.D. x 4.50" long centerless ground Pyrex® tubing.
  • The focusing trap chambers accept centerless ground Pyrex® tubing, 6 mm O.D. x 4.50" long.

Environment

    Range ambient to 50°C, 5 to 95% relative humidity. Corrosive substances (gas, liquid, or solid) may affect the materials and components used in the IACEM system and should be avoided.

Power Requirements

    115VAC single phase (110 to 125), 50 / 60 Hz, 8A maximum.

Overload Protection

    Operator replaceable, rear panel mounted, fuses protect the IACEM system against overload.

Dimensions

W x H x D

300 x 340 x 500 mm
(12 x 13.5 x 20")

Weight

19 Kg
(42 lbs.)