IACEM 980

The IACEM 980 air sampler is designed to quantitatively collect and release organic compounds from air on solid sorbent cartridges, transferring analytes to a gas chromatograph (GC) for separation, identification, and measurement. The IACEM was developed to enhance chemical agent workplace monitoring at the newly proposed lower airborne exposure levels.

The IACEM 980's dual collection tubes operate in alternate cycles providing continuous, uninterrupted sample concentration throughout the survey period. Each is linked in tandem with a focusing trap, enabling efficient chromatographic separation of agents from interferences. The focusing traps can be connected to capillary columns of dissimilar phases.

The result of this duplication is a blending of parallel systems into an integrated unit that promises sensitivity through increased sample collection efficiency and state-of-the-art chromatographic resolution. Selectivity is improved over current configurations through sophisticated comparisons of multiple-column retention data. The alternating collection scheme and duplicate columns increase speed of analysis through faster turnaround times.

Advanced Features:

Speed: New analytical results are available as often as every four minutes providing near real-time (NRT) performance. The cycle time is determined by the GC's oven program and cool-down time.

Selectivity: When configured with multiple dissimilar columns, same-cycle confirmation is realized which greatly reduces false positives.

Continuous: Two subsystems operate in a push-pull manner, 180 degrees out of phase, such that one sample is collected while the other is being analyzed. The result is uninterrupted sampling, providing maximum protection of personnel.

Sensitivity: Time-proven and patented construction and focusing techniques allow detection to 0.2 TWA (sub-part-per trillion) levels. High sampling flow rates (>1.0 lpm) permit the collection of quantities of analyte necessary for accurate and reproducible GC detection and quantification.

Flexibility: The IACEM may be used with GC's containing one, two, or four columns in combination with one or two detectors. The IACEM will operate effectively with new GC's or retrofit onto older GC models.

Product Specifications

Heated Zone-Setpoint

Valve Compartment: Ambient to 200°C
Transfer Line(s): Ambient to 260°C
Sample Tube Desorb: Ambient to 350°C
Sample Tube Idle: Ambient to 99°C
Focusing Trap Desorb: Ambient to 375°C
Focusing Trap Idle: Ambient to 99°C

Heating Accuracy

All heated zones ± 2% of range minimum

Heating Rates

Sample Tube: 1000°C/min. typical
Focusing Trap: 900°C/min. typical

Timed Events

A total of five sequence positions are designated: Ready (Idle), Tube Dry, Tube Heat, Trap Heat, and GC Time. Each event has a time (0 to 99.9 minutes in 0.1 minute increments) or off status (infinite).

Sampling Flow

  • Integral Porter® mass flow controller may be set from 0.00 to 2.55 slpm flow rate.

  • Accuracy & linearity: ±1% of full scale

  • Temperature Coefficient: 0.1%/°C of full scale

  • Pressure Coefficient: 0.1%/atmosphere

  • Repeatability: within 0.2% full scale at any constant temperature within operating temperature range

  • Response time: 1.2 seconds

Cycle Counter

Range is 1-7999 cycles, or off (infinite). System can be programmed to process a set number of samples, and then return to ready (idle) status.

Carrier Gas/Valving

  • Dual 8-port 1/8" stainless steel Valco® switching valves and electric actuators.

  • Sample transport tubing: sample path consists of 1/8" & 1/16" stainless steel tubing, silanized glass, and deactivated fused silica.

  • Transfer line: provision for two 1/4" O.D. heated jacket lines, one 36" heated jacket is supplied, internal surface is 0.25 mm I.D. fused silica.

  • Column flow: connections are provided for focusing trap desorb/column flow to be controlled by the EPC output from Agilent GCs. An optional flow control module is available for use if EPC is unavailable.

  • Glass tubing requirements: the sample collect chambers accept 6 mm and 10 mm (necked down to 6 mm) O.D. x 4.50" long centerless ground Pyrex® tubing.

  • The focusing trap chambers accept centerless ground Pyrex® tubing, 6 mm O.D. x 4.50" long.

Environment

Range ambient to 50°C, 5 to 95% relative humidity. Corrosive substances (gas, liquid, or solid) may affect the materials and components used in the IACEM system and should be avoided.

Power Requirements

115VAC single phase (110 to 125), 50 / 60 Hz, 8A maximum.

Overload Protection

Operator replaceable, rear panel mounted, fuses protect the IACEM system against overload.

Dimensions

W x H x D: 300 x 340 x 500 mm (12 x 13.5 x 20")
Weight: 19 Kg (42 lbs.)